Product type
|
SK-500II
|
|
Maximum injection rate
|
1500ml/h (Different injectors have different maximum rate.)
|
|
Range of flow rate
|
5ml injector: 0.1-100ml/h
10ml injector: 0.1-200ml/h
20ml injector: 0.1-400ml/h
30ml injector: 0.1-600ml/h
50ml injector: 0.1-1500ml/h
|
|
Injection increment
|
0.1ml/h
|
|
Bolus rate
|
5ml injector: 100ml/h
10ml injector: 200ml/h
20ml injector: 400ml/h
30ml injector: 400-600ml/h
50ml injector: 400-1500ml/h
|
|
Injection mode selection
|
1. Rate mode 2. Time mode 3. Body Weight mode
|
|
Rate mode
|
Speed: 0.1-1500ml/h (maximum value depends on the injector specifications)
|
|
Time mode
|
Time:1~2000 minutes
Accumulated injection volume: 0.1-999.9ml
|
|
Body Weight mode
|
Weight: 0.1~300.0kg
Drug: 0.1~999.9(mg)
Volume: 0.1~999.9ml
Dose: 0.1~9999.9(restricted to unit, drug, solution, weight and injector specification).
Unit: mg/kg/h, ug/kg/min
|
|
KVO rate
|
Adjustable range: 0.1ml/h~5.0ml/h (begin the KVO function under the occlusion situation) ; pressing STOP key to cancel the KVO function when it is unnecessary in use.
|
|
IP grade
|
IP21
|
|
Injection pressure
|
The maximum pressure is 0.3Mpa, Occlusion alarm pressure value is 40kPa~160kPa when pressure value is minimum, injection rate is 25ml/h, the longest alarm time is 10 minutes. When pressure value is minimum, injection rate is 5ml/h, the longest alarm time is 10 minutes.
|
|
Preset volume
|
0.1ml-9999.9ml
|
|
Preset time
|
0~99 hours, 0~59 minutes
|
|
Accumulated injection volume
|
0.1ml-9999.9ml
|
|
Accuracy
|
±3%
|
|
Power supply
|
AC 100~220V,50/60Hz
|
|
Battery
|
Rechargeable lithium polymer battery, 7.4 V, 1600mAh.
|
|
Maximum power consumption
|
25VA, running more than 2 hours at the rate of 25ml/h after being fully charged.
|
|
Battery charge
|
When syringe pump is connected to AC power, the battery will automatically recharge. (about 8~14 hours to recharge fully.)
|
|
Fuses
|
T2AL250V~
|
|
Displayed information
|
Flow rate; accumulated injection volume; injector specification; battery capacity; bed No.; AC power indicator, operation indication, etc.
|
|
State indication information
|
Stop, inject, bolus, KVO (stop flashing, the other indicators flash by turns.)
|
|
Alarm information
|
near completion, Over, occlusion, low battery, abnormal 1 (communications failure), abnormal 2 (pump stuck), abnormal 3 (wrong parameter), volume limited alarm, no AC power
|
|
Maximum size of the outer shell
|
288×130×122 mm(length×width×height),
|
|
Maximum weight
|
<3.0kg
|
|
Classification
|
Class?, Type BF
|
|
Shell material
|
ABS plastic
|
|
Operating conditions
|
Environment temperature 5?~40?, air pressure 80kPa ~106kPa,relative humidity ≤80%
|
|
Storage conditions
|
Environment temperature -40?~55?, air pressure 50kPa ~106kPa,relative humidity ≤95%
|
|
Applicable syringes
|
5ml?10 ml?20 ml?30ml?50 ml syringes
|
|
Electrical safety
|
In accordance with the stipulation of GB9706.1-1995
|
|
Applied standards
|
According to the stipulation of YZB/Yue 001-2008
|