Dimension
|
120×140×190 mm
|
|
Accumulated infusion volume
|
1.0~9999.0 ml
|
|
Maximum weight
|
< 3.0 kg
|
Infusion accuracy
|
±5%
|
|
Displayed information
|
Infusion rate, preset volume, accumulated infusion volume, battery capacity, AC power indicator, infusion completion, occlusion, air bubbles, bed No., pressure value, flow rate.
|
Alarm function
|
Over?occlusion?air bubbles?low battery?installation error of infusion pipe ?machine malfunction ?empty
|
|
Fuse
|
T 2AL 250V~
|
KVO rate
|
1.0 ml/h-5.0 ml/h
|
|
Maximum power consumption
|
25VA, running more than 4 hours after being charged
|
Selectable modes
|
Rate mode, Time mode, Weight mode
|
|
Infusion rate range
|
1.0-2000 ml/h
|
Rate mode
|
1.0~2000.0 ml/h
|
|
Bolus
|
Rate:≤1000.0 ml/h
Bolus: 1000.0 ml/h
|
Time mode
|
Volume limit: 0.1~999.9ml Time: 1~2000 minutes
|
|
Operation conditions
|
Environment temperature 5?~40?, air pressure 86kPa~106kPa, relative humidity ≤80%
|
Body Weight mode
|
Drug limit: 1.0~999.9 mg, Volume limit: 0.1~999.9 ml. Weight: 0.1~300.0 kg, Dose: 0.1~9999.9
Selectable unit: mg/kg/h, ug/kg/min
|
|
Electrical Safety
|
Comply with EN60601-1 for medical equipment and EN60601-2-24
|
Battery
|
Rechargeable lithium polymer battery, 7.4V, 1600 mA
|
|
Applicable infusion sets
|
All standard IV sets
|
Storage
|
Environment temperature-20~55?, air pressure 50~106kPa, relative humidity ≤95%
|