Maximum size of outer shell
|
120x140x190mm
|
|
Accumulated infusion quantity display
|
1~9999ml
|
|
Maximum weight
|
<2.5 kg
|
Accuracy
|
±5%
|
|
Power supply
(city electricity)
|
AC 220V, 50/60Hz
|
KVO rate
|
1ml/h, keep vein open (KVO) rate
|
|
displayed information
|
Flow rate,volume limit,accumulated volume,power indicator light,bed No.,air,occlusion,empty.
|
Alarm function
|
Infusion completion,occlusion,air bubble,low battery,control abnormal,no AC power supply, installation error.
|
|
Electrical safety
|
In according with the stipulation of GB9706.1-199.5
|
|
Electromagnetic wave protection
|
In according with the stipulation of IEC60601-1-2 (1993 edition).
|
|
Fuse
|
220V2A ×2, 12V 2A ×2
|
Infusion increment
|
1ml/h
|
|
Maximum power consumption
|
25VA, running more than two hours after being charged
|
Applicable infusion sets
|
All standard infusion sets( use " double dove" to test)
|
|
Infusion rate range
|
1-600ml/h
|
Feasible preset infusion volume
|
1~9999ml
|
|
Fast infusion volume
|
1.0ml (Fast infusion speed: 600ml/h)
|
Battery
|
Rechargeable lithium battery, 7.4V, 1650mAh
|
|
Environment condition:
|
environment temperature 5~40?,pressure 86~106kPa, relative humidity 20%~80%
|
Storage
|
Environment temperature
–15~50?, pressure 80~106kPa, relative humidity 20%~80%
|